Journal: Journal of Applied Clinical Medical Physics
Article Title: Optimization of the rounded leaf offset table in modeling the multileaf collimator leaf edge in a commercial treatment planning system
doi: 10.1120/jacmp.v15i6.4899
Figure Lengend Snippet: The default Pinnacle RLO table for the 80‐leaf Elekta MLC is shown in blue, the red and green curves are the modified RLO tables for a dosimetric‐based leaf calibration, and the purple curve is the modified RLO table for a light field‐based calibration.
Article Snippet: The graph of the default table in the Pinnacle software for the Elekta 80‐leaf MLC is shown in the blue curve of Fig. . For Pinnacle version 9.2, there is only one RLO table for all photon energies of a given machine, and only one table for both sets of leaves.